TECHNICAL SPECIFICATIONS
CALAJECT Type CJ2
Control unit
Handpiece
Stand
Length
95 mm
Foot switch cable: 2.2 m
200 mm (incl. barrel)
Handpiece cord: 1.7 m
Width
120 mm
Ø 12 mm
Ø 60 mm
Height
115 mm
34 mm
Weight
750 g
50 g
410 g
Nominal voltage
7.2 V, 1.2Ah
Battery (Lithium-ion)
5 hours on each charge
Charging time
Approx. 3 hours
Dental Cartridge
1.7/1.8 ml standard cartridges
Dental Needles
Standard M6 and 7/32”
Charging:
12V-DC/1ADC
PROGRAM 3
Recommended for regional nerve block analgesia
• Activate the footswitch and hold it down throughout the injection.
CALAJECT™ begins slowly, 0.006 ml/sec. By releasing/re-activating
the foot control in one swift movement, the injection speed will
increase gradually over the next 5 seconds to high speed
(approx. 0.04 ml/sec). Hereafter, high speed at every stop/start.
• Aspirates automatically whenever the foot switch is released.
The small suck-back will also prevent after-dripping from the
needle.
R
R
R
PROGRAM 2
Recommended for infiltration analgesia
• Activate the foot switch and hold it down throughout the injection.
The sequence begins with 10 seconds slow speed (0.006 ml/sec)
followed by a gradual increase over 5 seconds to medium
injection speed 0.03 ml/sec.
• When the injection has been stopped and is reactivated again, the
same sequence will repeat: First 10 seconds slowly with gradual
increase to medium speed. No option of speed change in Program 2.
• Aspirates automatically whenever the foot switch is released.
The small suck-back will also prevent after-dripping from the
needle.
R
R
R
CHARGER:
100-240v - 50-60hZ/280 - 140mA
ENVIRONMENT: Temperature:
Operation at 10 to 35°C / 50 to 95°F.
Storage and transportation at -20 to 50°C / -4 to 122°F.
Altitude:
Operation in 0-3000 m.
Storage and transportation: no restrictions.
Humidity:
Operation, storage and transportation: 10-95%.
Classification:
Council directive 93/42/EEC, Class IIa. Internally powered equipment, Class IIa battery charger, type b
applied part, ipx 7 foot control, suitable for continuous operation.
Standards:
EN60601-1
Disposal:
Separate collection of electronic waste.
RD_brugervejledning_IFU-CALAJECT-USA-2019.01-VER 7_090119.indd 7