Remote Commands
R&S
®
FPL1000
887
User Manual 1178.3370.02 ─ 09
Remote commands exclusive to marker demodulation:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
....................................887
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
..........................................887
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
............................................888
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
........................................... 888
.....................................................................................888
.................................................................................. 889
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
<State>
This command turns continuous demodulation of the signal at the marker position in
the frequency domain on and off.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC2:MARK3:FUNC:DEM:CONT ON
Switches on the continuous demodulation.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
<Duration>
This command defines for how long the the signal at the marker position is demodula-
ted.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
<m>
Parameters:
<Duration>
Range:
10 ms to 1000 s
*RST:
Marker demodulation = OFF
Default unit: S
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters