Remote Commands
R&S
®
FPL1000
885
User Manual 1178.3370.02 ─ 09
................................................................................ 885
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
............................................................885
CALCulate<n>:MARKer<m>:COUNt:RESolution
...............................................................886
CALCulate<n>:MARKer<m>:COUNt
<State>
This command turns the frequency counter at the marker position on and off.
The frequency counter works for one marker only. If you perform a frequency count
with another marker, the R&S
FPL1000 deactivates the frequency count of the first
marker.
The frequency counter is not available if the tracking generator is active (see
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation:
See
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the frequency at the marker position.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Setting Basic Spectrum RF Measurement Parameters