Remote Commands
R&S
®
FPL1000
672
User Manual 1178.3370.02 ─ 09
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:ABS
<Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see
"Relative limit line functions"
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STOP:FUNCtion
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
Chapter 11.6.7.10, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:FUNCtion
<Function>
This command enables the use of a function when defining the relative limit for a SEM
range.
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Function>
OFF | MAX
Defines the function to be used to determine the relative limit
line stop value
Configuring and Performing Measurements