Remote Commands
R&S
®
FPL1000
652
User Manual 1178.3370.02 ─ 09
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage:
Query only
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe
This command turns the relative limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<ch>
1..n
Alternate channel number
Parameters:
<State>
Manual operation:
See
CALCulate<n>:LIMit<li>:ACPower[:STATe]
<State>
This command turns the limit check for ACLR measurements on and off.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
11.6.4.6
Performing an ACLR Measurement
The following commands are required to perform an ACLR measurement:
●
CALC:MARK:FUNC:POW:SEL ACP
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
●
Configuring and Performing Measurements