![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 381](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989381.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
381
User Manual 1178.3370.02 ─ 09
Remote commands:
"Example: Obtaining a Marker Peak List"
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STATe
TRAC? LIST
, see
........................................................................................382
.............................................................................................382
Peak List State
Activates/deactivates the marker peak list. If activated, the peak list is displayed and
the peaks are indicated in the trace display.
For each listed peak the frequency/time ("X-value") and level ("Y-Value") values are
given.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STATe
Sort Mode
Defines whether the peak list is sorted according to the x-values or y-values. In either
case the values are sorted in ascending order.
Marker Usage