The Spectrum Application (RF Measurements)
R&S
®
FPL1000
295
User Manual 1178.3370.02 ─ 09
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
Reference Value
Defines an offset for the position of the reference line.
This setting can be used to shift the reference line and thus the normalized trace, simi-
lar to the
defined in the "Amplitude" settings shifts the ref-
erence level
in the display
.
Shifting the normalized trace is useful, for example, to reflect an attenuation or gain
caused by the measured DUT. If you then zoom into the diagram around the normal-
ized trace, the measured trace still remains fully visible.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
Recall Cal. Settings
Restores the settings that were used during source calibration. This can be useful if
instrument settings were changed after calibration (e.g. center frequency, frequency
deviation, reference level, etc.).
Remote command:
Save as Trd Factor
Uses the normalized measurement data to generate a transducer factor. The trace
data is converted to a transducer with unit dB and stored in a file with the specified
name and the suffix
.trd
under
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\trd
. The frequency points are allo-
cated in equidistant steps between start and stop frequency.
The generated transducer factor can be further adapted using the "Transducer" func-
tion in the [Setup] menu.
Note:
Note that the
normalized
measurement data is used, not the
reference
trace!
Thus, if you store the normalized trace directly after calibration, without changing any
settings, the transducer factor will be 0
dB for the entire span (by definition of the nor-
malized trace).
Remote command:
[SENSe:]CORRection:TRANsducer:GENerate
Power Sweep Settings
Access
: Toolbar > "Generator Config"
Or
: [Meas] > "Power Sweep" > "Generator Config"
Or
: [Freq] > "Generator Config"
The power sweep signal is currently only available in the Spectrum application, and
only if the optional Internal Generator R&S
FPL1-B9 is installed.
Receiving Data Input and Providing Data Output