Remote Commands
R&S
®
FPL1000
780
User Manual 1178.3370.02 ─ 09
Example:
DISP:TRAC:Y 110dB
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again (for all
traces).
Suffix:
<n>
.
<t>
irrelevant
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
<Mode>
This command selects the type of scaling of the y-axis (for all traces).
When the display update during remote control is off, this command has no immediate
effect.
Suffix:
<n>
.
<w>
subwindow
<t>
irrelevant
Parameters:
<Mode>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S
FPL1000 adjusts the scaling of the y-axis accordingly.
For measurements with the optional tracking generator, the command defines the posi-
tion of the reference line.
Suffix:
<n>
.
<t>
irrelevant
Setting Basic Spectrum RF Measurement Parameters