Remote control commands
R&S
®
NRP18S-xx
61
User Manual 1178.3686.02 ─ 07
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRP18S-xx sets the last used
length.
Example for
REAL,32
format:
#6768000....<binary float values>....
Example for
REAL,64
format:
#71536000....<binary float values>....
<data[,length]>
<ASCii[,0 to 12]>
ASCii
List of comma separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for
ASCii,4
format:
1.2938e-06, -4.7269e-11, ...
*RST:
ASCii,0
8.7
Configuring the measurement modes
In the following, the settings needed for configuring a measurement mode are descri-
bed.
Further information:
●
Chapter 8.8, "Configuring basic measurement parameters"
●
Chapter 8.5, "Controlling the measurement"
Contents:
Continuous average measurement
.........................................................................61
...................................................................................65
............................................................................................66
................................................................................................ 68
8.7.1
Continuous average measurement
The continuous average mode measures the signal average power asynchronously
within definable time intervals (sampling windows). The aperture (width of the sampling
windows) can be defined.
Reducing noise and zero offset
The smoothing filter can reduce result fluctuations caused by modulation. But activat-
ing it increases the inherent noise of the sensor by approx. 20
%., so do not activate if
it unless required.
Configuring the measurement modes