Remote Commands
R&S
®
FPL1000
806
User Manual 1178.3370.02 ─ 11
10.8.7.1
Configuring Standard Traces
Useful commands for trace configuration described elsewhere
●
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Remote commands exclusive to trace configuration
DISPlay[:WINDow<n>]:TRACe<t>:MODE
.........................................................................806
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE:HCONtinuous
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
............................................ 808
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
........................... 808
........................................................................................ 809
...................................................................................809
.......................................................................................... 810
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
............................................................... 810
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
..................................................... 811
DISPlay[:WINDow<n>]:TRACe<t>:MODE
<Mode>
This command selects the trace mode. If necessary, the selected trace is also activa-
ted.
In case of max hold, min hold or average trace mode, you can set the number of single
measurements with
. Note that synchronization to the end of
the measurement is possible only in single sweep mode.
Suffix:
<n>
.
<t>
Parameters:
<Mode>
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps. The "Sweep/Aver-
age Count" determines the number of averaging procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
FPL1000 saves the sweep result in the trace
memory only if the new value is greater than the previous one.
Setting Basic Spectrum RF Measurement Parameters