Remote Commands
R&S
®
FPL1000
712
User Manual 1178.3370.02 ─ 11
Chapter 10.6.1, "Performing Mea-
10.6.11.4
Retrieving Results
The following commands retrieve the results of the harmonic distortion measurement.
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
........................................ 712
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST
...................................................712
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
This command queries the total harmonic distortion of the signal.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Query parameters:
<Result>
TOTal
Return values:
<DistortionPct>
<DistortionDb>
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST
This command queries the position of the harmonics.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<Harmonics>
Returns one value for every harmonic.
The first value is the absolute power of the first harmonic. The
unit is variable.
The other values are power levels relative to the first harmonic.
The unit for these is dB.
Configuring and Performing Measurements