![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 631](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987631.webp)
Remote Commands
R&S
®
FPL1000
631
User Manual 1178.3370.02 ─ 11
Suffix:
<ch>
.
1..n
Alternate channel number
Parameters:
<Spacing>
Range:
100 Hz to 2000 MHz
*RST:
40 kHz (ALT1), 60 kHz (ALT2), 80 kHz (ALT3), ...
Default unit: Hz
Manual operation:
See
[SENSe:]POWer:ACHannel:SPACing:CHANnel<ch>
<Spacing>
This command defines the distance between transmission channels.
If you set the channel spacing for a transmission channel, the R&S
FPL1000 sets the
spacing of the lower transmission channels to the same value, but not the other way
round. The command works hierarchically: to set a distance between the 2nd and 3rd
and 3rd and 4th channel, you have to set the spacing between the 2nd and 3rd chan-
nel first.
Suffix:
<ch>
.
1..n
Tx channel number
Parameters:
<Spacing>
Range:
14 kHz to 2000 MHz
*RST:
20 kHz
Default unit: Hz
Manual operation:
See
[SENSe:]POWer:ACHannel:TXCHannel:COUNt
<Number>
This command defines the number of transmission channels.
The command works for measurements in the frequency domain.
Parameters:
<Number>
Range:
1 to 18
*RST:
1
Manual operation:
See
10.6.4.3
Defining Weighting Filters
The following commands define weighting filters for ACLR measurements.
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
........................................................ 632
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
............................................................... 632
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
................................................. 632
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>
.................................................. 632
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
....................................................... 633
Configuring and Performing Measurements