Remote Commands
R&S
®
FPL1000
633
User Manual 1178.3370.02 ─ 11
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
<State>
This command turns the weighting filter for the adjacent channel on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer[:STATe][:ALL]
<State>
This command turns the weighting filters for all channels on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
<State>
This command turns the weighting filter for an alternate channel on and off.
Suffix:
<ch>
.
1..n
Alternate channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
<State>
This command turns the weighting filter for a transmission channel on and off.
Suffix:
<ch>
.
1..n
Tx channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
10.6.4.4
Selecting the Reference Channel
The following commands define the reference channel for relative ACLR measure-
ments.
Configuring and Performing Measurements