General Instrument Setup
R&S
®
FPL1000
528
User Manual 1178.3370.02 ─ 11
........................................................................................... 528
..........................................................................................528
NONE
Uses the current RF signal at the input, i.e. no calibration signal (default).
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
Calibration Frequency RF
Uses the internal calibration signal as the RF input signal.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut:PULSed:CFRequency
Frequency ← Calibration Frequency RF
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration (max. 64
MHz).
Calibration Frequency MW
Uses the microwave calibration signal as the RF input. This function is used to cali-
brate the YIG-filter on the microwave converter. The microwave calibration signal is
pulsed. The microwave calibration signal frequency is fixed at 1.5
GHz.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
Service Functions