The Spectrum Application (RF Measurements)
R&S
®
FPL1000
299
User Manual 1178.3370.02 ─ 11
Level Offset
Defines an offset to the output power of the internal generator.
Used to adapt the level display, for example to cable loss.
Remote command:
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
PS Frequency
Defines the frequency of the internal generator signal for a power sweep measure-
ment. This frequency corresponds to the center frequency of the RF measurement.
Remote command:
Frequency Offset
Shifts the frequency of the internal generator by the defined offset.
The default frequency of the internal generator is coupled to the (center) frequency of
the analyzer and can be set from 5
kHz to the maximum frequency of the
R&S
FPL1000.
Remote command:
SOURce<si>:INTernal:FREQuency:OFFSet
Level Start / Level Stop
Defines the start and stop levels of the output power of the internal generator. The
power increases linearly within this range. Only ascending values are allowed, that is:
the level start value must be lower than the level stop value.
The range is from -60
dBm to +10
dBm.
is defined, the "Level Start" / "Level Stop" values are indicated with the
offset applied.
Remote command:
Calibrate Transmission
Starts a transmission type measurement to determine a reference trace. This trace is
used to calculate the difference for the normalized values.
Remote command:
Normalization state
Switches the normalization of measurement results on or off. This function is only
available if the memory contains a reference trace, that is, after a calibration has been
performed.
Remote command:
Receiving Data Input and Providing Data Output