![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 791](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987791.webp)
Remote Commands
R&S
®
FPL1000
791
User Manual 1178.3370.02 ─ 11
SOURce<si>:INTernal:FREQuency
<Frequency>
Defines the frequency of the internal generator signal.
Suffix:
<si>
.
irrelevant
Parameters:
<Frequency>
Range:
5 kHz to 3 GHz
Increment: 0.1 Hz
*RST:
1 GHz
Default unit: HZ
Example:
SOUR:INT:FREQ 2 GHz
Manual operation:
See
SOURce<si>:POWer[:LEVel][:IMMediate][:AMPLitude]
<Amplitude>
Defines the output power of the internal generator.
Suffix:
<si>
.
irrelevant
Parameters:
<Amplitude>
Range:
-60 dBm to +10 dBm
Increment: 0.1 dB
*RST:
-20 dBm
Default unit: DBM
Example:
SOUR:POW -30dBm
Manual operation:
See
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
This command defines a level offset for the external generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the external generator can be taken into
account for the setting.
Suffix:
<si>
.
irrelevant
Parameters:
<Offset>
Range:
-200 dB to +200 dB
*RST:
0dB
Default unit: DB
Example:
//Define a level offset on the external generator
SOUR:POW:OFFS -10dB
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters