Remote Commands
R&S
®
FPL1000
868
User Manual 1178.3370.02 ─ 09
If the marker hasn't been active yet, the command first looks for the peak level in the
whole spectrogram and moves the marker vertically to the minimum level.
Suffix:
<n>
.
<m>
11.8.8.7
Fixed Reference Marker Settings
The following commands configure a fixed reference marker.
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
............................................. 868
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y
............................................. 869
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y:OFFSet
................................. 869
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe]
............................................... 869
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
This command moves the fixed reference marker to the peak power.
Suffix:
<n>
.
<m>
Example:
CALC:DELT:FUNC:FIX:RPO:MAX
Sets the reference point level for delta markers to the peak of
the selected trace.
Manual operation:
See
See
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
<RefPoint>
This command defines the horizontal position of the fixed delta marker reference point.
The coordinates of the reference may be anywhere in the diagram.
Suffix:
<n>
.
<m>
Parameters:
<RefPoint>
Numeric value that defines the horizontal position of the refer-
ence.
For frequency domain measurements, it is a frequency in Hz.
For time domain measurements, it is a point in time in s.
*RST:
Fixed Reference: OFF
Default unit: HZ
Example:
CALC:DELT:FUNC:FIX:RPO:X 128 MHz
Sets the frequency reference to 128 MHz.
Setting Basic Spectrum RF Measurement Parameters