Remote Commands
R&S
®
FPL1000
726
User Manual 1178.3370.02 ─ 09
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<TOI>
Maximum third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES:MAX?
Returns the maximum TOI.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult:MINimum?
This command queries the results for the minimum third order intercept point measure-
ment (see
Chapter 7.2.12.3, "TOI Results"
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<TOI>
Minimum third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES:MIN?
Returns the minimum TOI.
Usage:
Query only
Configuring and Performing Measurements