Remote Commands
R&S
®
FPL1000
628
User Manual 1178.3370.02 ─ 09
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sweeps) and waits for the
end.
Result: Averaging is performed over 40 sweeps.
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command controls the sweep mode for an individual channel setup.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see
Chapter 10.1.6, "Command Sequence and Syn-
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
0
Example:
INIT:CONT OFF
Switches the sweep mode to single sweep.
INIT:CONT ON
Switches the sweep mode to continuous sweep.
Configuring and Performing Measurements