![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 568](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985568.webp)
Network and Remote Operation
R&S
®
FPL1000
568
User Manual 1178.3370.02 ─ 09
Bit No.
Meaning
4
ALT2 UPPer FAIL
This bit is set if the limit is exceeded in the
upper 2nd alternate
channel.
5
ALT2 LOWer FAIL
This bit is set if the limit is exceeded in the
lower 2nd alternate
channel.
6
ALT3 … 11 LOWer/UPPer FAIL
This bit is set if the limit is exceeded in one of the lower or upper alternate channels 3 … 11.
7
CACLR FAIL
This bit is set if the CACLR limit is exceeded in one of the gap channels.
8-14
Unused
15
This bit is always 0.
STATus:QUEStionable:EXTended Register
The
STATus:QUEStionable:EXTended
register contains further status information
not covered by the other status registers of the R&S
FPL1000. A separate
EXTended
register exists for each active channel.
You can read out the register with
STATus:QUEStionable:EXTended:CONDition?
or
STATus:QUEStionable:EXTended[:EVENt]?
Table 10-11: Meaning of the bits used in the STATus:QUEStionable:EXTended register
Bit No.
Meaning
0
not used
1
INFO
This bit is set if a status message is available for the application.
Which type of message occurred is indicated in the
STATus:QUEStionable:EXTended:INFO
.
2 to 14
Unused
15
This bit is always 0.
STATus:QUEStionable:EXTended:INFO Register
The
STATus:QUEStionable:EXTended:INFO
register contains information on the
type of messages that occur during operation of the R&S
FPL1000. A separate
INFO
register exists for each active channel.
You can read out the register with
STATus:QUEStionable:EXTended:INFO:
or
STATus:QUEStionable:EXTended:INFO[:EVENt]?
. You can
query all messages that occur for a specific channel using the command