Data Management
The R&S
FPL1000 allows you to save and recall measurement settings. Measurement
data can be exported and imported for a later analysis, the graphical result display can
be stored to a file or printed. In addition, various application-specific data can be saved
and/or recalled.
General storage and import/export functions are available via the toolbar. Some special
storage functions are (also) available via softkeys or dialog boxes in the corresponding
menus, for example trace data export.
Restoring the Default Instrument Configuration (Preset)
...................................... 456
Storing and Recalling Instrument Settings and Measurement Data
........................................................................................467
Creating Screenshots of Current Measurement Results and Settings
.................................................................................... 485
8.1
Restoring the Default Instrument Configuration (Pre-
set)
When delivered, the R&S
FPL1000 has a default configuration. You can restore this
defined initial state at any time as a known starting point for measurements. This is
often recommendable as a first step in troubleshooting when unusual measurement
results arise.
Factory default configuration
The factory default configuration is selected such that the RF input is always protected
against overload, provided that the applied signal levels are in the allowed range for
the instrument.
Alternatively to the factory default settings, you can define user-specific recall settings
to be restored after a preset or reboot, see
"To recall settings automatically after preset
To restore the default instrument configuration for all channel setups at once
► Press the [PRESET] key.
After you use the [PRESET] function, the history of previous actions is deleted, i.e. any
actions performed previously cannot be undone or redone using the [UNDO/REDO]
keys.
Remote command:
*RST
or
SYSTem:PRESet
Restoring the Default Instrument Configuration (Preset)