The Spectrum Application (RF Measurements)
R&S
®
FPL1000
438
User Manual 1178.3370.02 ─ 09
already been zoomed graphically. In this case, the temporary changes to the display
are replaced by permanent changes to the measurement settings with the same effect.
Example:
Assume you have a spectrum display from a spurious emission measurement. You
graphically zoom into the area around a detected spur. If you now activate a measure-
ment zoom, the reference level, the center frequency, the frequency span, and the
scaling settings are adapted so that the results of the measurement now indicate only
the formerly zoomed area around the detected spur.
Single Zoom Versus Multiple Zoom
...................................................................... 438
................................................................................ 441
7.12.1
Single Zoom Versus Multiple Zoom
Two different (graphical) zoom modes are available: single zoom and multiple zoom. A
single zoom replaces the current diagram by a new diagram which displays an
enlarged extract of the trace. This function can be used repetitively until the required
details are visible. In multiple zoom mode, you can enlarge up to four different areas of
the trace simultaneously. An overview window indicates the zoom areas in the original
trace, while the zoomed trace areas are displayed in individual windows. The zoom
areas can be moved and resized any time. The zoom area that corresponds to the indi-
vidual zoom display is indicated in the lower right corner, between the scrollbars.
Figure 7-63: Single zoom
Zoomed Displays