The Spectrum Application (RF Measurements)
R&S
®
FPL1000
125
User Manual 1178.3370.02 ─ 09
General power sweep settings are described in
...............................................................................125
Compression Measurement State
Enables or disables the compression point measurement.
Remote command:
CONFigure:CMEasurement[:STATe]
Reference
Determines the reference point for the linear trace on which the deviation or compres-
sion calculation is based.
By default, the reference point is automatically set to the start level of the power sweep
(see
on page 297). In manual mode, define the reference
point as required.
For details, see
Remote command:
CONFigure:CMEasurement:REFerence[:MODE]
CONFigure:CMEasurement:REFerence:VALue
Compression Points
Selects which compression points to calculate. For the user-defined compression
point, define the distance from the linear trace to the reference trace to base the calcu-
lation on. For details, see
Chapter 7.2.3.1, "About the Power Sweep Compression
Measurements and Results