MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS 

The Romed Infrared Forehead Thermometer is intended for use in the electromagnetic environments specified 

below. The customer or the user of the Romed Infrared Forehead Thermometer should assure that it is used in 

such an environment.

 

Emission test

 

Compliance

 

Electromagnetic environment-guidance

 

RF emissions

 

CISPR 11

 

Group 1

 

The Romed Infrared Forehead Thermometer uses RF energy 

only for its internal function. Therefore, its RF emissions are very 

low and are not likely to cause any interference in nearby 

electronic equipment.

 

RF emissions

 

CISPR 11

 

Class B

 

The Romed Infrared Forehead Thermometer is suitable for use 

in all establishments, including domestic establishments and 

those directly connected to the public low-voltage power supply 

network that supplies buildings used for domestic purposes.

 

Harmonic emissions

 

IEC 61000-3-2

 

Class A

 

Voltage fluctuations/flicker 

emissions

 

IEC 61000-3-3

 

Complies

 

MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY 

The Romed Infrared Forehead Thermometer  is intended for use in the electromagnetic environment specified 

below. The customer or the user of The Romed Infrared Forehead Thermometer  should assure that it is used 

in such an environment.

 

Immunity test

 

IEC 60601 test level

 

Compliance level

 

Electromagnetic environment -

guidance

 

Electrostatic 

discharge (ESD) 

IEC 61000-4-2

 

±8 kV contact

 

± 2 kV, ± 4 kV,

 

± 8 kV, ± 15 kV air

 

±8 kV contact

 

± 2 kV, ± 4 kV,

 

± 8 kV, ± 15 kV air

 

Floors should be wood, concrete 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%.

 

Electrostatic 

transient / burst 

IEC 61000-4-4

 

±2kV for power supply lines

 

±1kV for input/output lines

 

Not applicable

 

Not applicable

 

Surge

 

IEC 61000-4-5

 

±1kV differential mode

 

±2kV common mode

 

Not applicable

 

Not applicable

 

Voltage dips, short 

interruptions and 

voltage variations 

on power supply 

input lines

 

IEC 61000-4-11

 

< 5% UT (>95% dip

 

in UT) for 5 sec

 

Not applicable

 

Not applicable

 

Power frequency 

(50/60 Hz)

 

magnetic field

 

IEC 61000-4-8

 

30 A/m

 

30 A/m

 

Power frequency magnetic fields 

should be at levels characteristic 

of a typical location in a typical 

commercial or hospital 

environment.

 

NOTE      UT  is the AC mains voltage prior to application of the test level.

 

10

www.biomark.lv

Содержание THERM-IR

Страница 1: ...read this Instruction Manual before using the unit in order to use it safely it safely and correctly The instruction manual should be well kept for your reference at any time H O L L A N D H O L L A N D USER MANUAL INFRARED FOREHEAD THERMOMETER THERM IR w w w b i o m a r k l v ...

Страница 2: ... the device This device can be applied to all people No special training is needed but the user shoul read the manual carefully before use 1 Foreword 2 Device Description 3 Instructions for Use 4 Specifications 5 Cleaning and Disinfection 6 Maintenance 7 Trouble Shooting 8 Disposal 9 Electromagnetic Compatibility INDEX 1 2 4 7 8 8 8 9 9 w w w b i o m a r k l v ...

Страница 3: ...ying the device Please do not clean or perform maintenance on the device while in use Please wait for 30 minutes before use if the environment has a sudden change e g from under sunlight to an air conditioned room or the accuracy may be influenced Degraded sensors and electrodes or loosened electrodes can degrade performance or cause other problems DO NOT open the battery cover around any source o...

Страница 4: ...ing the body temperature by measuring the heat radiation from the forehead Probe Type BF Applied part Serial number Batch code Catalogue number CE Mark Date of manufacture Manufacturer Keep dry Keep away from sunlight Temperature limit Collect separately from other household waste IP classification Refer to user manual Consult instructions for use Caution 2 w w w b i o m a r k l v ...

Страница 5: ...Button 4 Thermosensor 5 Distance Sensor 6 Battery Cover 7 Body Mode Indicator 8 Object Mode Indicator 9 Temperature reading distance indicator 10 C F CM 11 Memory Indicator 12 Memory Times Remaining time 13 Battery Charge 14 Measure Distance 15 Batteries 3 w w w b i o m a r k l v ...

Страница 6: ...r leave the device for 30 minutes before operation if it is moved from different environment and make sure the device always works in the correct range temperature 3 Position thermometer up to 3 cm away from the centre of the forehead just between the eyebrows If the eyebrow area is covered with hair sweat or dirt please clean the area beforehand to improve the reading accuracy It is important to ...

Страница 7: ...p 3 FEVER GUIDANCE FEATURE Fever guidance helps you to better understand the meaning of the temperature with the colour indicated on the display The screen displays green is normal yellow when the temperature is low fever and a red alert for a possibly high fever 10 audible beeps indicate temperatures above 38 5 C 101 3 F to alert the user that the patient may have a fever Backlight Colour Tempera...

Страница 8: ...f In the memory view interface press and hold the measurement button for approx 5 seconds to display CLR and delete the all memories CHANGING MEASUREMENT MODE In the ON status press the power button to switch between the object mode and the body temperature mode Object mode is able to measure the surrounding environment or object temperature To ensure the accuracy of measurement please do not cond...

Страница 9: ... 0 1 C 0 1 F Measurement Accuracy 0 2 C for range 35 0 C 42 0 C 0 3 C outside this temperature range at standard room temperature of 25 C 77 0 F Memory 30 recorded measurements Button 3 buttons Power Model Button Measurement Memory Button C F Button Alarm Sound when peak temperature reached Calibration No need for calibration before use To ensure the measurement accuracy please use and store the d...

Страница 10: ... The continuous use of a damaged unit may cause injury improper results or serious danger Store and use the device at a cool dry and ventilated environment Avoid approaching fire and heat sources or it will cause the battery to explode If you have any problems with this device such as setting up maintaining or using or in the case of an unexpected operation or event please contact Van Oostveen Med...

Страница 11: ...ns section 8 DISPOSAL DO NOT dispose the batteries in domestic waste Dispose of the batteries according to the local regulations dealing with the disposal of these special materials e g to the collecting points The device is made of a combination of plastic and stainless steel Be sure to dispose of it in accordance with local regulations as unsorted municipal waste You may recycle it at your local...

Страница 12: ...rehead Thermometer is intended for use in the electromagnetic environment specified below The customer or the user of The Romed Infrared Forehead Thermometer should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 8 kV contact 2 kV 4 kV 8 kV 15 kV air 8 kV contact 2 kV 4 ...

Страница 13: ... the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths form fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE...

Страница 14: ...ximum output of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the freq...

Отзывы: