MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
The Romed Infrared Forehead Thermometer is intended for use in the electromagnetic environments specified
below. The customer or the user of the Romed Infrared Forehead Thermometer should assure that it is used in
such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The Romed Infrared Forehead Thermometer uses RF energy
only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Romed Infrared Forehead Thermometer is suitable for use
in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
The Romed Infrared Forehead Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of The Romed Infrared Forehead Thermometer should assure that it is used
in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
±8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrostatic
transient / burst
IEC 61000-4-4
±2kV for power supply lines
±1kV for input/output lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1kV differential mode
±2kV common mode
Not applicable
Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5% UT (>95% dip
in UT) for 5 sec
Not applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the AC mains voltage prior to application of the test level.
10