Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
NOTE
: Before
use by a different individual, clean, disinfect, and check product for proper functioning.
Cleaning and Disinfecting
WARNING:
If using bleach, follow safety guidelines on bleach container.
CAUTION:
DO NOT:
• use abrasives such as steel wool or scouring pads.
• use cleaning products containing petroleum or organic solvents such as acetone, toluene, Methyl
Ethyl Ketone (MEK), naphtha, dry cleaning fluids, adhesive removers.
• use caustic, automatic dishwasher detergents.
• allow water or cleaning solution to enter the product.
• expose the product to ozone gas or ultraviolet light cleaning methods.
Cover Care:
Before cleaning and disinfecting, remove hook fastener pieces from bottom of cover and
turn cover inside out. Wash with like colors.
To Clean the Cover:
Machine wash in warm water (40°C) with mild detergent,
gentle cycle, or wipe clean with neutral detergent and warm water. Tumble dry
low.
To Disinfect the Cover
: Hand wash in warm water using
1 part household liquid bleach per 9 parts water. Rinse
thoroughly. Cover can also be machine washed in hot water
(60°C), gentle cycle, to disinfect. Tumble dry low.
To Clean the Cushion by Hand:
Hand wash in warm (40°C) water. Use
mild liquid hand soap, hand dishwashing liquid, laundry detergent, or
multipurpose detergent that is used for general cleaning. Use a soft,
plastic bristle brush, sponge, or washcloth to gently scrub all cushion surfaces
.
Rinse with clean
water. Air dry.
DO NOT
machine dry.
DO NOT
expose to direct sunlight.
To Disinfect the Cushion by Hand:
Repeat cleaning
instructions using 1 part household liquid bleach per 9 parts
water.
DO NOT
autoclave.
Or
To Disinfect the Cushion by Machine:
Deflate the cushion by opening
the valve and rolling up the cushion, starting from the corner opposite
of the valve. When the cushion is deflated, close the valve. Place the
cushion in the washing bag. Close the washing bag. Machine wash in
hot (60°C) water or use the sanitize cycle. Machine settings: Use mild detergent, gentle cycle with no
spin. Air Dry.
DO NOT
machine dry.
DO NOT
expose to direct sunlight.
DO NOT
autoclave.
To Sterilize the Cushion:
Follow Cleaning and Disinfection guidelines. ROHO products are
not packaged in a sterile condition nor are they intended or required to be sterilized prior to
usage. High temperatures accelerate aging and may cause damage to the product assembly. If
institution protocol demands sterilization, open the valve and use the lowest temperature possible
for the shortest time possible. Gas sterilization is preferable to steam autoclave, but neither is
recommended. In any process using pressure, ensure all valves are open to prevent damage to the
product. ROHO, Inc. highly discourages sterilization methods using temperatures greater than 160°F
(71°C) for polyurethane products.
40C
60C
1:9
40C
or
+
40C
or
1:9
+
40C
60C