Performance Test
R&S
ZVB
1145.1078.82 1.2
E–4
Test Equipment and Accessories
Item Type
of
equipment
Recommended characteristics or
features
Recommended
model
R&S
Order No.
Application
1
Spectrum
analyzer
a) Counter mode:
Min. resolution: 100 Hz
Max. rel. frequency deviation: 10
–6
b) Linearity
Max. departure from linearity (2 ):
0.06 dB
c) Frequency response:
50 MHz to 4 GHz: < 1 dB
4 GHz to 8 GHz: < 1.5 dB
8 GHz to 20 GHz: < 2 dB
R&S FSU 26
1129.9003.26 Frequency uncertainty
Harmonics
Output linearity
Input linearity
2
Power sensor
N, 50 . 300 kHz to 8 GHz
RSS error < 0.1 dB
VSWR < 1.2
R&S NRP–Z51 1138.0005.02 Max. output power
Accuracy of output power
Power measurement
uncertainty on R&S ZVB4/8
2
Power sensor
3.5mm, 50 . 10 MHz to 20 GHz
RSS error < 0.1 dB
VSWR < 1.25
R&S NRP–Z55 1138.2008.02 Max. output power
Accuracy of output power
Power measurement
uncertainty on
R&S ZVB14/20
2
Power meter
300 kHz to 20 GHz
R&S NRVD
+ Sensor
R&S ZV–Z51
0857.8008.02
0857.9004.02
Max output power
Accuracy of output power
Power measurement
uncertainty
4
Calibration kit
N, 50 . 300 kHz to 8 GHz.
R&S ZV–Z21
1085.7099.02 Input linearity
Matching port 1 to port 4
Input noise level
on R&S ZVB4/8
4
Calibration kit
3.5 mm, 50 . 10 MHz to 20 GHz.
R&S ZV–Z32
1134.4293.02 Input linearity
Matching port 1 to port 4
Input noise level
on R&S ZVB14/20
5
Signal
generator
300 kHz to 20 GHz,
Power = –40 dBm to 10 dBm
R&S SMR27
with Option
R&S SMR–B15
R&S SML01
1104.0002.27
1104.4989.02
1090.3000.11
Power measurement
uncertainty
6
Power splitter
N, 50 Ohm,
eq
< 0.05 (50 MHz to 8 GHz)
Output tracking < 0.15 dB
Weinschel
1870A
–
Power measurement
uncertainty
6
Power splitter
3.5mm, 50 Ohm,
Output tracking < 0.25 dB
Weinschel
1593
–
Power measurement
uncertainty
7
Test cable
N(male) – N(male), approx. 1.5 m
Frequency uncertainty
Harmonics
Matching port 1 to port 4
Power measurement
uncertainty
Input linearity
on R&S ZVB4/8
Содержание ZVB 14
Страница 4: ......
Страница 10: ......
Страница 56: ......
Страница 58: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 4 ...
Страница 66: ......
Страница 154: ...Table of Contents Software Update Installation of Options R S ZVB 1145 1078 82 I 4 2 E 4 ...
Страница 158: ...5 2 ...
Страница 160: ...1145 1078 82 5 2 E 4 Spare Parts List Mechanical Drawings ...
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 183: ......
Страница 184: ......
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 188: ......
Страница 189: ...1145 1078 82 5 19 E 4 Block Circuit Diagram ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......