Glossary
R&S
®
ZVA
101
Quick Start Guide 1145.1090.62 ─ 10
Trace:
A trace is a set of data points that can be displayed together on the screen. The
trace settings specify the mathematical operations used to obtain traces from the collec-
ted data. They complement the definitions of the Channel menu. Each trace is assigned
to a channel. The channel settings apply to all traces assigned to the channel.
Trace point:
Point on the screen which is an element of the displayed trace. The trace
points for ratios and wave quantities can be derived from the entire set of measurement
points using different detector settings.
Transmission tracking error:
Frequency-dependent variation of the ratio of the trans-
mitted wave to the reference wave at a test port when an ideal transmission coefficient
(= 1) is measured. The transmission tracking error can be corrected by means of a trans-
mission normalization or one of the more sophisticated calibration methods.
TRL:
A calibration type using a through and a line standard plus a possibly unknown
reflect standard. The calibration range is limited. Like TNA and TRM, TRL is especially
useful for DUTs in planar line technology.
TRM:
A calibration type using a through and a match standard plus a possibly unknown
reflect standard. Like TRL and TNA, TRM is especially useful for DUTs in planar line
technology.
W
Window:
Rectangular portion of the screen showing all diagram areas of a particular
setup. Windows are limited by a blue frame with several icons. The analyzer uses stand-
ard windows provided by the operating system.