Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
197
User Manual 1178.5989.02 ─ 06
9.7.3.6
Band Power Marker
The following commands control the marker for band power measurements.
Using Markers
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:AOFF
..................................................... 197
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE
.................................................... 197
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?
..................................................197
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN
..................................................... 198
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe]
...................................................198
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:AOFF
Removes all band power markers in the specified window.
Suffix:
<n>
.
<m>
irrelevant
Example:
CALC:MARK:FUNC:BPOW:AOFF
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE
<Mode>
This command selects the way the results for a band power marker are displayed.
(Note: relative power results are only availabe for delta markers,
see .
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
Suffix:
<n>
.
<m>
Parameters:
<Mode>
POWer
Result is displayed as an absolute power. The power unit
depends on the
setting.
DENSity
Result is displayed as a density in dBm/Hz.
*RST:
POWer
Example:
CALC:MARK4:FUNC:BPOW:MODE DENS
Configures marker 4 to show the measurement results in
dBm/Hz.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
I/Q Analysis