4. Define the sweep range
5. Adjust the receiver and source settings (measurement bandwidth, source power)
6. Perform a calibration
3.4.1 Transmission S-Parameter Measurement
In a transmission measurement, the analyzer transmits a stimulus signal to the input
port of the device under test (DUT) and measures the transmitted wave at the DUT's
output port. The trace settings allow you to select the measured quantities and display
formats, depending on what you want to learn from the data. A minimum of two ana-
lyzer test ports are required for transmission measurements.
In the following example, the analyzer is set up for a two-port transmission measure-
ment. A frequency sweep range is selected, the instrument is calibrated and the mea-
surement result is analyzed using various display formats.
3.4.1.1
Connecting the Instrument for Transmission Measurements
In order to prepare a transmission measurement, you have to connect your DUT
(which for simplicity we assume to have appropriate connectors) in-between a pair of
analyzer test ports. It is recommended that you preset the R&S
ZNB/ZNBT in order to
start from a well-defined instrument state.
Performing Measurements