GUI Reference
R&S
®
ZNB/ZNBT
608
User Manual 1173.9163.02 ─ 38
Note:
In contrast to the R&S
ZNB the R&S
ZNBT doesn't have a built-in loudspeaker
(PC speaker). Hence alarm sounds are not audible unless the R&S
ZNBT is operated
via remote desktop.
Remote command:
SYSTem:SOUNd:ALARm[:STATe]
SYSTem:SOUNd:STATus[:STATe]
Use Default Tab for Hardkey
If the checkbox is selected (system default), a front panel key or a button in the hard-
key panel activates the first enabled tab of the associated softtool. Otherwise the last
used tab is activated.
For background information see
Chapter 5.1, "Hardkeys and Softtools"
Remote command:
n/a
Conductance in Embedding Networks
Changes the presentation of "capacitance C<i> in parallel with resistance R<i>" circuit
blocks in lumped de/embedding networks (see
Chapter 4.6.2.3, "Circuit Models for 2-
Chapter 4.6.2.4, "Circuit Models for 4-Port Net-
If active, the resistance R<i> is displayed and specified as conductance G<i> (=1/
R<i>).
Remote command:
n/a
Decimal Places
Defines the number of fractional digits for quantities with different physical units. The
settings affect entries and results, e.g. the values in the marker lists.
Note:
If your instrument is equipped with option R&S
ZNB/ZNBT-K19, 1 mHz Fre-
quency Resolution, set "Decimal Places" of unit "Hz" to
12
in order to utilize the high
frequency resolution.
Remote command:
n/a
Units Prefix
Sets the unit prefix for frequencies (Base unit: Hz) to Kilo (k), Mega (M), Giga (G) or
Tera (T) or lets the R&S
ZNB/ZNBT select the appropriate prefix (Auto = default set-
ting).
Remote command:
n/a
Reset Colors / Reset Dialogs / Reset Decimal Places / Reset Units Prefix
Resets the color settings ("System" > "Display" < "Config"), the dialog properties, the
"Decimal Places" and the "Units Prefix" settings. These settings are global and not
affected by an instrument preset.
Remote command:
n/a