Auto Power Reduction for Cal Unit
Sets the source power at all test ports to –10 dBm while an automatic calibration is
active. Applying this source power to the ports of the calibration unit ensures best
accuracy of the automatic calibration. The source powers are reset to their original val-
ues after the calibration is completed. The automatic power reduction can be deactiva-
ted in case that the test setup introduces a large attenuation.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:PREDuction[:STATe]
Auto Averaging
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to regular aver-
aging (see
on page 238) the number of calibration
sweeps is calculated automatically.
Remote command:
[SENSe:]CORRection:COLLect:AVERage
Multiple Cal in Calibration Wizard
Enables/disables multiple calibrations in the calibration wizard, which allows to set up
and perform multiple calibrations in one manual calibration procedure. See
ter 5.4.1.3, "Calibration Unit Wizard (MultiCal)"
tion Presetting Wizard (MultiCal)"
Chapter 5.4.1.8, "SMARTerCal Wizard
Chapter 5.4.1.10, "SMARTerCal Wizard (Man-
Remote command:
SYSTem:DISPlay:DIALogs:SETup:MCAL[:STATe]
Show Cal Kit Label
Enables/disables the "Calibration Info" dialog during manual calibration (see
Setup Softtool