GUI Reference
R&S
®
ZNB/ZNBT
499
User Manual 1173.9163.02 ─ 38
Time Gate
Enables or disables the time gate for the time domain and frequency domain traces.
"Gat" is displayed in the trace list while the time gate is active.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe
Axis Pair
"Start Stop" lets you define the time gate via its "Start" and "Stop", "Center Span" lets
you define it via its "Center" and "Span" value (in time). The analyzer generates a
warning if the (resulting) time span exceeds the unambiguous range which is given by
Δt = 1/Δf, where Δf is the "Freq Step Size". Simply reduce the time span until the warn-
ing disappears.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt
CALCulate<Chn>:FILTer[:GATE]:TIME:STOP
Show Range Lines
Displays or hides two red lines indicating the start and stop of the time gate in a time
domain diagram.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW
Bandpass / Notch
The filter type defines what happens to the data in the specific time region.
●
A "Bandpass" filter passes all information in the specified time region and rejects
everything else.
Trace Config Softtool