GUI Reference
R&S
®
ZNB/ZNBT
331
User Manual 1173.9163.02 ─ 38
Step 2: Save Characterization Data
Saves the characterization data to the calibration unit.
Figure 5-24: Save Characterization Data
Save File to SD Card
For some calibration units (e.g. the new models R&S
ZN-Z5x and R&S
ZN-Z15x),
characterization data can either be saved to the calibration unit's internal flash memory
or to an SD card inserted at the calibration unit. Activate this checkbox to save the
characterization data to the SD card.
Tip
: If the characterized CalUnit doesn't have an SD card slot, the checkbox is hidden.
If the CalUnit has an SD card slot but the SD card isn't accessible, the checkbox is
greyed out.
File Name/ Comment (Optional)
Selects a filename to reference the characterization data set in the "Characterize Cal
Unit" and "Calibration Unit" dialogs and a comment, to be written into the characteriza-
tion file. A filename is required before you can "Finish" the characterization wizard and
store the data.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:CKIT
Cal Softtool