GUI Reference
R&S
®
ZNB/ZNBT
257
User Manual 1173.9163.02 ─ 38
Sweep All Channels
Apply the sweep control settings to all channels in the active recall set. The number of
sweeps in a "Single" sweep sequence is equal to the selected number of "Sweeps"
times the number of channels. The sequence starts with the first sweep in channel no.
1.
Tip:
In remote control, it is possible to retrieve the results acquired in any of the
sweeps within a single sweep group.
Remote command:
INITiate<Ch>[:IMMediate]:SCOPe
Sweep Active Channel
Apply the sweep control settings to the active channel only. The number of sweeps in a
"Single" sweep sequence is equal to the number of "Sweeps" in the active channel.
The table lists all channels in the active recall set and allows you to define individual
numbers of sweeps for all channels. When a new channel is selected, the analyzer
uses its specific number of sweeps.
Remote command:
INITiate<Ch>[:IMMediate]:SCOPe
SINGle
5.4 Cal Softtool
The "Cal" softtool provides all functions related to system error calibration, scalar
power calibration, and "SMARTerCal".
Access:
CHANNEL – CAL hardkey
5.4.1 Start Cal Tab
The "Start Cal" tab provides access to all functions for automatic or manual calibration.
Calibration of the R&S
ZNB/ZNBT is a fully guided process.
Cal Softtool