Command Reference
R&S
®
ZNB/ZNBT
1189
User Manual 1173.9163.02 ─ 38
AR1 | AR2 | AR3 | AR4 | BR1 | BR2 | BR3 | BR4 |
CR1 | CR2 | CR3 | CR4 | DR1 | DR2 | DR3 | DR4 |
R1A | R1B | R1C | R1D | R2A | R2B | R2C | R2D |
R3A | R3B | R3C | R3D | R4A | R4B | R4C | R4D
Ratio of wave quantities b
1
/a
1
, b
1
/a
2
, ... , b
4
/a
4
, a
1
/b
1
,
a
1
/b
2
, ..., A
4
/B
4
(received waves to reference waves
or reference waves to received waves)
R1R2 | R1R3 | R1R4 | R2R1 | R2R3 | R2R4 | R3R1
| R3R2 | R3R4 | R4R1 | R4R2 | R4R3
Ratio of wave quantities a
1
/a
2
, a
1
/a
3
, ... , a
4
/a
3
(refer-
ence waves only)
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth
<Bandwidth>
Defines the bandfilter level, i.e. the minimum excursion for the bandpass and bandstop
peaks.
Tip:
Use
CALCulate<Chn>:MARKer<Mk>:BWIDth
to set the bandwidth and query
the results of a bandfilter search. Note the sign convention for input values.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is
ignored and may be set to any value because the bandfilter
search functions always use markers M1 to M4.
Parameters:
<Bandwidth>
Range:
-100 dB to 100 dB
Increment: 0.03 dB
*RST:
3 dB
Default unit: dB
Example:
See
CALCulate<Chn>:MARKer<Mk>:BWIDth
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DELTa:STATe
<Boolean>
Switches the delta mode for marker
<Mk>
on trace no.
<Chn>
on or off.
Note:
This command is the ZVR-compatible equivalent of
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters:
<Boolean>
ON | OFF - enable or disable the delta mode.
*RST:
OFF
CALCulate<Chn>:MARKer<Mk>:FUNCtion:TARGet
<SearchValue>
Defines the target value for the target search of marker no.
<Mk>
, which can be activa-
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
TARGet
.
Note:
This command is the ZVR-compatible equivalent of
.
R&S ZVR/ZVAB Compatible Commands