Command Reference
R&S
®
ZNB/ZNBT
1181
User Manual 1173.9163.02 ─ 38
Example:
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.
Manual operation:
See
TRIGger<Ch>[:SEQuence]:SOURce
<TrigSource>
Selects the source of the trigger events that the analyzer uses to start a measurement
sequence.
Suffix:
<Ch>
.
Channel number
Parameters:
<TrigSource>
IMMediate | EXTernal | MANual | MULTiple
IMMediate – free run measurement (untriggered).
EXTernal – trigger by external signal applied to the EXT TRIG IN
connector or pin 2 of the USER PORT on the rear panel.
MANual – trigger event generated by pressing the "Manual Trig-
ger" softkey.
MULTiple – multiple trigger mode, configured by
TRIGger<Ch>[:SEQuence]:MULTiple...
commands.
Example:
TRIG:SOUR MAN
Activate manual trigger mode. The analyzer starts the next mea-
surement sequence when the "Manual Trigger" button is
pressed.
Manual operation:
See
"FreeRun / External / Manual / Multiple Triggers"
7.3.20.2
TRIGger:CHANnel:AUXiliary...
The
TRIGger:CHANnel<Ch>:AUXiliary...
commands control the external trigger
output EXT TRIG OUT at the rear panel of the instrument. The output trigger signal
consists of TTL pulses of configurable duration and polarity which may be sent before
or after each sweep or measurement point.
Figure 7-1: Output trigger signal at EXT TRIG OUT
SCPI Command Reference