Command Reference
R&S
®
ZNB/ZNBT
1132
User Manual 1173.9163.02 ─ 38
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt>
Physical port number. This suffix is ignored; the setting is valid
for all ports.
Parameters:
<ReduceAt
SweepEnd>
OFF
(default): at sweep end the output power of the sweep start
is restored
ON
: at sweep end the output power of the driving port is reduced
as if the channel base power was set to its minimum possible
value
Example:
See
SOURce<Ch>:POWer<PhyPt>:SWEepend:MODE
REDuce:STATe OFF
is equivalent to
SWEepend:MODE AUTO
REDuce:STATe ON
is equivalent to
SWEepend:MODE REDuce
SOURce<Ch>:POWer<PhyPt>:REDuce:SDELay
<SettlingDelay>
Deprecated; superseded by
SOURce<Ch>:POWer<PhyPt>:SWEepend:SDELay
If power reduction at sweep end is enabled (
ON) this defines the settling delay at subsequent sweep starts, i.e.
the time between power-up and sweep start when a new sweep is requested.
This is a global setting.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt>
Physical port number. This suffix is ignored; the setting is valid
for all ports.
Parameters:
<SettlingDelay>
Settling delay
Default unit: s
Example:
See
SOURce<Ch>:POWer<PhyPt>:SWEepend:MODE
SOURce<Ch>:POWer<PhyPt>:STATe
<Boolean>
Turns the RF source power at a specified test port on or off.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer
SCPI Command Reference