Command Reference
R&S
®
ZNB/ZNBT
1032
User Manual 1173.9163.02 ─ 38
[SENSe<Ch>:]FREQuency:CONVersion
<ConversionMode>
Enables a frequency conversion measurement mode for channel
<Ch>
.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConversionMode>
FUNDamental | ARBitrary
FUNDamental - measurement of the fundamental signal (no fre-
quency-converting mode). This selection cancels all port-specific
frequency and power settings (see example).
ARBitrary - frequency conversion mode (arbitrary port frequen-
cies). This mode is automatically activated when a port-specific
frequency is defined.
MIXer scalar mixer mode. The mixer mode is configured using
the
[SENSe<Ch>:]FREQuency:CONVersion:MIXer...
commands.
*RST:
FUNDamental
Example:
See
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:
Manual operation:
See
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary
<Numerator>, <Denominator>,
<Offset>, <SweepType>
Unused command.
The R&S
ZNB/ZNBT doesn't allow to set a separate receiver frequency for
Frequency Conversion Measurements
. Use
CONVersion:ARBitrary:IFRequency
to define the port-specific frequency (range
or CW).
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary:PMETer<Pmtr>
<Numerator>,
<Denominator>, <Offset>, <SweepType>
Defines the receiver frequency of a power meter used for frequency-converting mea-
surements. The receiver frequency is either a range (for frequency sweeps) or a CW
frequency (for power, time and CW Mode sweeps). The receiver frequency is valid for
all ports.
<Numerator>, <Denominator> and <Offset> are parameters of the frequency formula.
The receiver frequency f
r
is calculated according to
f
r
= <Numerator>/<Denominator> * f
b
+ <Offset>
where f
b
represents the channel base frequency (parameter
SWEep
). For parameters
CW
or
FIXed
, f
b
= 0.
SCPI Command Reference