Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
131
User Manual 1179.3315.02 ─ 01
DISPlay[:WINDow<n>]:TRACe<t>:MODE
<Mode>
This command selects the trace mode. If necessary, the selected trace is also activa-
ted.
Suffix:
<n>
.
<t>
Parameters:
<Mode>
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps. The "Sweep/Aver-
age Count" determines the number of averaging procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
VSE saves the sweep result in the trace mem-
ory only if the new value is greater than the previous one.
MINHold
The minimum value is determined from several measurements
and displayed. The R&S
VSE saves the sweep result in the
trace memory only if the new value is lower than the previous
one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
BLANk
Hides the selected trace.
*RST:
Trace 1: WRITe, Trace 2-6: BLANk
Example:
INIT:CONT OFF
Switching to single sweep mode.
SWE:COUN 16
Sets the number of measurements to 16.
DISP:TRAC3:MODE WRIT
Selects clear/write mode for trace 3.
INIT;*WAI
Starts the measurement and waits for the end of the measure-
ment.
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
<Detector>
Sets and queries the detector for the selected result display.
Suffix:
<n>
.
Configuring Standard Traces