Preparing for use
R&S
®
TSMExxDC
12
Manual 4901.1040.02 ─ 04
●
R&S TSME-Z1 has an output power of about 30 W, sufficient for up to two
products, one TSME6 and one TSME30/44DC.
●
R&S TSMA6-Z1 has an output power of 105 W and is suitable for multiple
TSMx products.
1. Ensure that the required ratings listed in the data sheet are matched.
2. Connect the round connector to DC IN.
3. Insert the AC power plug into a power outlet with ground contact.
3.3
Connecting to RF In
To connect RF cable (SMA connector)
► Connect the RF cable with SMA plug to the RF In (
≤
6 GHz, bypass). See
and
.
The SMA connector is sensitive to mechanical stress. Use the following
handling precautions.
●
Always use a torque wrench and mount the cable end with 60 Ncm.
●
Do not stack adapters directly at the SMA connector. If you need to use
adapters, then always use a specific adapter cable (order no.
4900.1700.00).
To connect RF cable (K connector)
►
NOTICE!
Using cables with SMA plugs may damage the K connector (bad RF
transition from plug to connector).
Connect the RF cable with K plug to the following RF in connectors.
● Use only cables with K plugs for the connection to the K connector.
● Use the port saver (R&S TDC292JACK, order no. 3637.8397.02) for the K
connector.
)
● RF In 24 GHz to 27.5 GHz
● RF In 27 GHz to 30 GHz
Connecting to RF In