Installation and Commissioning
R&S
®
TS7124AS
24
Instructions Handbook 1179.2948.02 ─ 01
6
Installation and Commissioning
The following activities are restricted to the
Familiarize yourself with residual risks and potentially dangerous situations.
See
Chapter 2.3, "Potentially Danger-
Execute these activities in the same order as given in this handbook:
...................................................................................24
............................................................................................26
................................................................................... 31
Connecting to Control and Power
........................................................................... 33
...........................................................................................37
Defining Zones of Restricted Access
...................................................................... 38
....................................................................................... 40
6.1
Choosing the Operating Site
Only use the chamber indoors. The chamber casing is not waterproof.
Select an operating site that provides safe conditions for mounting and operating the
chamber.
Ensure the following:
●
Only trained personnel can access the operating site, with restrictions as described
in
Chapter 6.7, "Defining Zones of Restricted Access"
●
The room has an even floor and sufficient bearing capacity.
●
The operating site leaves sufficient room to open the door without obstruction and
to access:
–
Chamber, especially the area behind the open door
–
Connectors on the front and rear side
–
Mounting brackets
–
Panic button or power plug, see
"Prerequisites for power connection"
●
The environmental conditions such as ambient temperature and humidity match
the values in the data sheet.
●
The operating site is at an altitude of maximum 2000
m above sea level.
●
The environment has pollution degree
2, where only nonconductive contamination
occurs. Occasionally, temporary conductivity caused by condensation is to be
expected.
●
The electromagnetic compatibility (EMC) class of the chamber is class
A.
Choosing the Operating Site