![Rohde & Schwarz SMM100A Скачать руководство пользователя страница 342](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939342.webp)
Configuring the RF signal
R&S
®
SMM100A
342
User Manual 1179.1341.02 ─ 05
[:SOURce<hw>]:LFOutput:FREQuency:STARt
[:SOURce<hw>]:LFOutput:FREQuency:STOP
Center Frequency
In "RF Frequency Sweep" mode, sets the RF center frequency.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
Remote command:
[:SOURce<hw>]:FREQuency:CENTer
Span
In "RF Frequency Sweep" mode, sets the span of the frequency sweep range.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
Remote command:
Spacing
In "RF and LF Frequency Sweep" mode, "Spacing " selects the mode for calculating
the frequency interval, which increases or decreases the current frequency at each
step.
To determine the step size, select the parameter
"Linear"
Takes the frequency value entered as absolute value in Hz.
"Logarithmic"
Takes the value entered as a logarithmic value, i.e. as a constant
fraction of the current frequency in %.
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
Step Linear/Step Logarithmic
In "RF/LF Frequency Sweep" mode, sets the step width for the individual frequency
sweep steps. The value is added at each sweep step to the current frequency.
Depending on the current
, you can enter either an absolute or logarithmic step
width.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
"Step Linear"
The step width is a constant value in Hz.
"Step Logarithmic"
The step width is determined logarithmically in %, i.e. as a constant
fraction of the current frequency.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
Varying the RF signal in list or sweep mode