Remote Control Commands
R&S
®
SMF100A
531
Operating Manual 1167.2319.02 ─ 12
Parameters:
<PowCntrlPeak>
0 | 1 | OFF | ON
*RST:
0
Example:
POW:SPC:PEAK ON
uses the measured peak power for power control.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:SELect
<PowCntrlSelect>
Defines the currently selected sensor to be used for power control.
Parameters:
<PowCntrlSelect>
SENS1 | SENS2 | SENS3 | SENS4
*RST:
SENS1
Example:
POW:SPC:SEL SENS2
selects the sensor connected to a second USB interface for
power control.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:STATe
<PowCntrlState>
Activates power control using the selected sensor. The control loop periodically adjusts
the generator output. After switching off, the running loop is completed.
Parameters:
<PowCntrlState>
0 | 1 | OFF | ON
*RST:
0
Example:
POW:SPC:STAT ON
activates power control.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:TARGet
<PowCntrlTarget>
Sets the nominal level expected at the input of the sensor. To define the unit of the
power value, use command
Parameters:
<PowCntrlTarget>
float
Range:
-50 to 30
Increment: 0.01
*RST:
-10
Example:
SENS:UNIT dBm
selects unit dBm for setting the target level value.
POW:SPC:TARG -10
sets -10 dBm target level.
SOURce Subsystem