Remote Control Commands
R&S
®
SMF100A
435
Operating Manual 1167.2319.02 ─ 12
:SENSe<ch>[:POWer]:ZERO
The command activates the autozero function. Zeroing is required in regular interval (at
least once a day) and if the temperature has varied more than about 5 °C, if the sensor
has been replaced or if measurements of signals with very low power are to be per-
formed. The RF power source must be switched off or disconnected from the sensor
before starting the autozero function.
Example:
SENS:ZERO
activates autozero function.
Usage:
Event
Manual operation:
See
:SENSe<ch>:UNIT[:POWer]
<Power>
The command selects the unit used for result query with command READ. The power
sensor provides the measured value in Watt. In which unit the measured value is
returned is selected here and might be either Watt, dBm or dBuV.
Parameters:
<Power>
DBM | DBUV | WATT
*RST:
DBM
Example:
SENS2:UNIT DBM
selects unit dBm for the measured value returned by command
READ
.
READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation:
See
7.12.6 TRACe Subsystem
:TRACe<ch>[:POWer]:SWEep:COLor
..............................................................................437
.............................................................................. 437
:TRACe<ch>[:POWer]:SWEep:DATA:POINts?
..................................................................437
:TRACe<ch>[:POWer]:SWEep:DATA:XVALues?
...............................................................438
:TRACe<ch>[:POWer]:SWEep:DATA:YVALues?
...............................................................438
...............................................................................438
:TRACe[:POWer]:SWEep:MEASurement:FULLscreen:DISPlay:ANNotation[:STATe]
:TRACe[:POWer]:SWEep:MEASurement:GATE:DISPlay:ANNotation[:STATe]
:TRACe[:POWer]:SWEep:MEASurement:MARKer:DISPlay:ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:AVERage?
......................................440
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:HREFerence?
................................. 440
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:LREFerence?
.................................. 440
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:MAXimum?
..................................... 440
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:MINimum?
...................................... 440
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:PULSe:BASE?
................................ 440
Power Sensor Measurement Subsystems