Remote Control Commands
R&S
®
SMF100A
530
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Resolution>
R01 | R1
*RST:
R01
Example:
POW:RES R01
sets the resolution for the level settings to 0.01 dB.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:CRANge
<PowCntrlCRange>
Defines the capture range of the power control system.
Within the range:
Target Level +/- Catch Range
the power control locks and tries to achieve the target level. Readings outside the
range are not considered.
Parameters:
<PowCntrlCRange> float
Range:
0 to 50
Increment: 0.01
*RST:
30
Default unit: dB
Example:
POW:SPC:CRAN 15
sets the capture range to +/- 15 dB.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:DELay
<PowCntrlDelay>
Defines a waiting period between the level adjustment of the generator and the next
measurement of the power sensor.
Parameters:
<PowCntrlDelay>
integer
Range:
0 to 1000
*RST:
0
Example:
:POW:SPC:DEL 2 ms
the sensor starts the next reading 2 ms after the level adjust-
ment.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:PEAK
<PowCntrlPeak>
Activates power control by means of the peak power values, provided the power sen-
sor supports this function.
SOURce Subsystem