Instrument Function
R&S
®
SMF100A
182
Operating Manual 1167.2319.02 ─ 12
The display has been disabled for security reasons, see
.
●
The display is disabled when list mode is running, see
Remote command:
Note:
The SCPI command
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
sets the level of the "Level" display, that is the level containing offset.
Limit - RF Level
Sets an upper limit for the RF output power.
You can use it to protect your DUT from damage due to high input power. If you enter
an RF level above this value, the instrument limits the output power to this specified
value, and generates a warning message.
However, the level indication in the status bar is not affected.
Note:
The limit value is always effective, regardless of whether you work with "NRP-Z
Power Control" or not.
The value is not affected by an instrument preset (PRESET key),
*RST
and the "Save/
Recall" function. It is influenced only by the
and the factory value is
equal to maximum level.
Remote command:
[:SOURce<hw>]:POWer:LIMit[:AMPLitude]
Offset (Level)
Sets the level offset relative to the RF level.
The level offset of a downstream instrument (for example an attenuator or amplifier) is
entered.
The entry does not change the value of the RF level at the RF output. It only changes
the RF level displayed in the display header. The value of the RF level in the header
corresponds to the level at the output of the downstream instrument.
Remote command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
Power Resolution
Selects the resolution for the level settings.
Remote-control command
:SOUR:POW:RES 0.01
Remote command:
[:SOURce<hw>]:POWer:RESolution
5.4.2.2
User Variation
If the level is set using the rotary knob, the step width is defined in the "User Variation"
section.
RF Level