Remote Control Commands
R&S
®
SMF100A
487
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Mode>
AUTO | MANual | STEP
AUTO
Performs a complete sweep cycle from the start to the end value
when a trigger event occurs.
The dwell time deterrmines the time period for the signal to
switch to the next step.
MANual
Performs a single sweep step when a manual trigger event
occurs.
The trigger system is not active. You can trigger each frequency
step of the sweep individually with the command
SOURce<hw>]:LFOutput:FREQuency:MANual
mode, use the rotary knob for switching to the next step.
With each step, the frequency increases by the value specified
with the command
[:SOURce<hw>]:LFOutput:SWEep[:
LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
respectively. A frequency value, entered with
With manual control, the frequency increases or decreases
(depending on the direction of the rotary encoder) by the value
specified under
SOUR:LFO:SWE:FREQ:STEP:LIN
(linear spac-
ing) or
...:STEP:LOG
(logarithmic spacing).
STEP
Each trigger triggers one sweep step only. The frequency
increases by the value entered with
LFOutput:SWEep[:FREQuency]:STEP[:LINear]
SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:
LOGarithmic
*RST:
AUTO
Example:
LFO:SWE:MODE AUTO
selects Auto mode.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:POINts
<Points>
Determines the number of steps for the LF frequency sweep within the sweep range.
This parameter always applies to the currently set sweep spacing and correlates with
the step size as follows:
●
for linear sweeps and f
STARt
< f
STOP
freq_points = (f
SPAN
/ step_lin) + 1
with f
SPAN
= f
STOP
- f
STARt
To determine the step size, use the command
.
SOURce Subsystem