Instrument Function
R&S
®
SMF100A
279
Operating Manual 1167.2319.02 ─ 12
Mode
Selects the LF frequency sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Generates a continuously repeating sweep signal immediately after
activating the sweep mode.
The sweep steps are performed atomatically, controlled by the dwell
time, see
Example:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR AUTO
SOUR:LFO:FREQ:MODE SWE
"Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
To trigger the sweep, use "Execute Single Sweep" button, or the cor-
responding remote control commands, for example
*TRG
.
Example:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR SING
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:SWE:FREQ:EXEC
"Step"
Generates the sweep signal step-by-step, manually triggered.
To perform the sweep steps, enter the frequency value under
Example:
SOUR:LFO:SWE:FREQ:MODE MAN
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:SWE:FREQ:SPAC LIN
SOUR:LFO:SWE:FREQ:STEP:LIN 1E34
SOUR:LFO:FREQ:MAN 12 kHz
The value entered with command
SOUR:LFO:SWE:FREQ:STEP:LIN|LOG
sets the step width.
The value entered with command
SOUR:LFO:FREQ:MAN
has no
effect, the command only sets the next sweep step. In remote control
only a step-by-step sweep from start to stop frequency is possible.
Modulation Generator, Pulse Generator and LF Output