Remote Control Commands
R&S
®
SMF100A
392
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
:DISPlay[:WINDow][:POWer]:SWEep:GRID:STATe
<State>
Indacates a grid in the diagram.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
DISP:SWE:GRID:STAT OFF
deactivates the indication of a grid in the diagram area.
Manual operation:
See
7.12.3 INITiate Command
:INITiate<ch>[:POWer]:CONTinuous
<Continuous>
The command switches the local state of the continuous power measurement by the
R&S NRP-Zxx power sensors on and off. Switching the local state off enhances the
measurement performance during remote control
The remote measurement is triggered by the READ query (command
on page 392) which also provides the measurement results. The local state
is not influenced by this command, measurements results can be retrieved with local
state on or off.
Parameters:
<Continuous>
0 | 1 | OFF | ON
*RST:
OFF
Example:
INIT:CONT ON
switches local state of continuous power measurement on.
Manual operation:
See
7.12.4 READ Subsystem
:READ<ch>[:POWer]?
The command triggers the measurement with power sensors and provides the power
measurement result of the selected power sensor. The value is provided with the unit
set with command
SENSe:UNIT[:POWer]
.
For certain power sensors, e.g. R&S NRP-Z81, two values are returned, first the value
for the average level and - separated by a comma - the peak level
Note:
The local state is not influenced by this command, measurements results can be
retrieved with local state on or off. For long measurement times it is recommended to
use a SRQ (MAV bit) for command synchronization.
Power Sensor Measurement Subsystems