Preparing for Use
R&S
®
SMC100A
19
Operating Manual 1411.4060.32 ─ 04
2.2 Rear Panel Tour
This section gives an overview of connectors on the rear panel of the instrument. Each
connector is briefly described and a reference is given to the chapters containing detailed
information. For technical data of the connectors refer to the data sheet.
Fig. 2-2: Rear panel view
2.2.1 Description of the Connectors
AC SUPPLY AND POWER SWITCH
When the R&S SMC is connected to the AC supply, it automatically sets itself to the
correct range for the applied voltage (range: see type label). There is no need to set the
voltage manually or change fuses.
The power switch can be set to two positions:
●
0
The instrument is disconnected from the mains.
●
I
The instrument is power-supplied. It is either ready for operation (STANDBY) or in
operating mode, depending on the position of the ON/STANDBY switch on the instru-
ment front.
chapter 2.3.3, "Connecting the Instrument to the AC Sup-
FUSES
The R&S SMC is fully fused by two fuses IEC60127-T3.15H/250 V.
The fuses are accommodated in the fuse holders next to the power connector. Use only
fuses of the mentioned type.